Precautions for Use 1
Precautions for Use
Contraindications
Do not use the 920M PLUS in an MRI environment.
Do not use the 920M PLUS in situations where alarms are required. The
device has no alarms.
Warnings
Explosion hazard: Do not use the 920M PLUS in an explosive atmosphere.
The 920M PLUS is intended only as an adjunct in patient assessment. It
must be used in conjunction with other methods of assessing clinical signs
and symptoms.
Use only Respironics compatible sensors, which are manufactured to meet
the accuracy specifications for Respironics pulse oximeters. Using other
manufacturers’ sensors may cause improper pulse oximeter performance.
As with all medical equipment, carefully route patient cabling to reduce the
possibility of patient entanglement or strangulation.
Check the pulse oximeter sensor application site frequently to determine the
positioning of the sensor and the circulation and skin sensitivity of the
patient. Each patient's sensitivity to Respironics sensors may vary
depending on medical status or skin condition.
Do not use adhesive tape strips if the patient exhibits an allergic reaction.
Do not stretch the adhesive tape while applying the pulse oximeter sensor.
This may cause inaccurate readings or skin blisters.
General operation of the 920M PLUS might be affected by the use of an
electrosurgical unit (ESU).
Do not use a damaged sensor.
Cautions
Federal law (USA) restricts this device to sale by or on the order of a
physician.
Before use, carefully read the package insert provided with the sensors.
The 920M PLUS is intended for spot-checking or monitoring by an
attending healthcare professional. Because the 920M PLUS has no SpO
2
alarms, international labeling requirements (EN 865) dictate it be labeled
not for continuous monitoring, as indicated by the symbol.
The 920M PLUS Pulse Oximeter is not an apnea monitor.
Verify that all visible indicators illuminate during the startup (initialization)
sequence. If any indicator is not lit, do not use the 920M PLUS. Contact
Respironics Customer Support for repair or replacement.