Respironics 920M Manuel de l'opérateur Page 36

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 89
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 35
33
Pulse Quality
display does not
blink green.
Inadequate pulse signal
or
the sensor is not
correctly positioned.
Reposition the sensor.
The sensor is restricting
blood circulation at the
sensor site.
Remove the restriction
to increase blood
circulation at the
sensor site or
relocate
the sensor.
Excessive ambient light. Reduce ambient light.
Excessive patient
motion.
Reduce patient
motion.
The patient is wearing
nail polish or artificial
nails.
Remove nail polish or
artificial nails.
Performance
degradation from:
arterial catheter
blood pressure cuff
infusion line
Reduce or eliminate
the source.
The Pulse Quality
display is blinking
red and the SpO
2
and/or Pulse Rate
displays are
dashes.
Inadequate signal at
sensor site.
Assess the patient.
Reposition sensor or
select an alternate
sensor site.
Inadequate pulse signal
due to excessive motion.
Reduce patient
motion. Reposition or
relocate the sensor.
Sensor failure. Replace the sensor.
Note: If these solutions do not correct the problem with your device,
please contact Respironics Technical Service at (800) 345-6443 or
(724) 387-4000.
Problem Possible Cause Possible Solution
Vue de la page 35
1 2 ... 31 32 33 34 35 36 37 38 39 40 41 ... 88 89

Commentaires sur ces manuels

Pas de commentaire