14 User Manual
al a r m pr i o r i t y au D i b l e
in D i c a t o r
vi S u a l in D i c a t o r S De v i c e ac t i o n uS e r ac t i o n
Low
Pressure
Alarm
High • • • • • Red ashing button;
“Low Pressure”
Operates This could be caused by an excessive
leak or blockage or a device malfunction.
Press the Alarm Silence/Indicator button
to silence the alarm. Remove your mask.
Check for the following: dirty inlet lters,
blocked air intake, excessive leak in the
patient circuit. If the alarm continues,
contact your home care provider.
Low Speed High • • • • • Red ashing button;
“Low Blower Speed”
Operates This may be caused by a malfunctioning
device. Press the Alarm Silence/Indicator
button to silence the alarm. Remove
your mask Remove power from the
device. Restore power. If the alarm
continues, contact your home care
provider for service.
High
Pressure
High • • • • • Red ashing button;
“High Pressure”
Operates;
If the alarm
continues for
10 seconds, the
alarm escalates
to a Device
Inoperative alarm
This may be caused by a malfunctioning
device. Press the Alarm Silence/Indicator
button to silence the alarm. Remove
your mask Remove power from the
device. Restore power. If the alarm
continues, contact your home care
provider for service.
Low
Pressure
Support
High • • • • • Red ashing button;
“Low Pressure
Support”
Operates This could be caused by an excessive
leak or blockage or a device malfunction.
Press the Alarm Silence/Indicator button
to silence the alarm. Remove your mask.
Check for the following: dirty inlet lters,
blocked air intake, excessive leak in the
patient circuit. If the alarm continues,
contact your home care provider.
Apnea High • • • • • Red ashing button;
“Apnea”
Operates This alarm is generated when an apnea
event occurs during therapy. Press the
Alarm Silence/Indicator button to silence
the alarm. Report the alarm to your
home care provider. Continue using your
device.
Low Minute
Ventilation
High • • • • • Red ashing button;
“Low Minute Vent”
Operates This alarm occurs when the calculated
minute ventilation is less than or equal
to the alarm setting. Press the Alarm
Silence/Indicator button to silence the
alarm. Report the alarm to your home
care provider. Continue using your
device.
Patient
Disconnect
High • • • • • Red ashing button;
“Patient Disconnect”
Operates This alarm occurs when the patient
circuit is disconnected or has a large
leak. Press the Alarm Silence/Indicator
button to silence the alarm. Reconnect
the patient circuit or x the leak. If the
alarm continues, contact your home care
provider for service.